Durr Hygoclave 50
Hygoclave 50. Hygiene is in our DNA.
With Hygoclave 50, Dürr Dental offers dental practices a professional Class B solution that combines impressive performance with a well thought-out operating concept – for maximum efficiency and exceptional user friendliness designed to cope effortlessly with tough day-to-day working environments.
Robust stainless steel chamber
Stainless steel and plastic are used across the board in the sterilisation process to prevent the the risk of highly corrosive demineralised water dissolving ions from the materials. Chemical reactions that could impair the sterilisation process are suppressed as a result.
High-performance process control
Regardless of the size of the dental practice, the required throughput or the structural conditions of the building: when it comes to fresh water supply, every practice has to go its own way. Hygoclave 50 has proved itself to be a flexible partner capable of offering the perfect solution for any constellation.
1. Manual fresh water supply
2. Co1mbination with a water treatment system
3. Integrated pump
Display
With the intuitive operation of its graphical interface, the high-resolution 5″ touch screen delivers comfortable standalone operation of all functions. Colour-coded icons on the home screen make for incredibly easy navigation and offer fast access to key functions and the integrated user management system. There, the most recently executed sterilisation and test cycles can be conveniently started via quick access or programme starts can be freely programmed.
Take a look at the Hygoclave 50 / 50 Plus Brochure
KEY FACTS
Powerful
High-performance sterilisation and drying system
Intuitive
Intuitive operating concept thanks to high-resolution 5″ touch screen, automatic door locking mechanism and flexible fresh water supply
Safe
Extra universal programme for additional safety even when sterilising complex hollow bodies
Connected
Everything in view and documented thanks to connection to the practice network and dental documentation programmes
Technical data Hygoclave 50 Hygoclave 50 Plus
Supply voltage 220 V–60 Hz , 230 V–50 /
60 Hz, 240 V–50 Hz
220 V–60 Hz , 230 V–50 /
60 Hz , 240 V–50 Hz
Maximum current consumption at 230 V 8.9 A 9.6 A
Maximum power consumption at 230 V 2050 W 2200 W
External dimensions (W x H x D) 493 x 484 x 642 mm 493 x 484 x 642 mm
Weight 57 kg 61 kg
Chamber dimensions (Ø x D) 250 x 370 mm 250 x 450 mm
Chamber volume 17 l 22 l
Chamber material Stainless steel Stainless steel
Fresh water container 5.2 l 5.2 l
Maximum load for standard cycles Type B Hygoclave 50 Hygoclave 50 Plus
Unpackaged, bulky load 5.5 kg 6.5 kg
Packaged, bulky load 3.5 kg 4 kg
Porous load 1.5 kg 2 kg
Maximum load for B flash cycle Hygoclave 50 Hygoclave 50 Plus
Packaged, bulky load 0.5 kg 0.7 kg
Porous load 0.2 kg 0.3 kg
Programs Hygoclave 50 Hygoclave 50 Plus
134°C B Standard (hold time 5.5 min) 45 min 48 min
134°C B Standard Plus (hold time 5.5 min) 49 min 51 min
134°C B Prions (hold time 20 min) 60 min 62 min
134°C B Flash (hold time 5.5 min) 33 min 34 min
121°C B Standard (hold time 20.5 min) 58 min 60 min
*Average values determined at maximum load, including pre-heating phase and drying. Standards: EN 13060 / EN 61010-1 / EN 61326 / EN 61010-2-40
Reviews(0)